Acumed Hexalob Screw various sizes: REF Numbers: : 3070-27008 2.7mm x 8mm NL Hexalobe Screw; 3070-27009 2.7mm x 9mm NL Hexalobe Screw; 3070-27010 2.7mm x 10mm NL Hexalobe Screw; 3070-27011 2.7mm x 11mm NL Hexalobe Screw; 3070-27012 2.7mm x 12mm NL Hexalobe Screw; 3070-27013 2.7mm x 13mm NL Hexalobe Screw; 3070-27014 2.7mm x 14mm NL Hexalobe Screw; 3070-27015 2.7mm x 15mm NL Hexalobe Screw; 3070-27016 2.7mm x 16mm NL Hexalobe Screw; 3070-27017 2.7mm x 17mm NL Hexalobe Screw; 3070-27018 2.7mm x 18mm NL Hexalobe Screw; 3076-30010 3.0mm x 10mm NL Hexalobe Screw; 3076-30012 3.0mm x 12mm NL Hexalobe Screw; 3076-30014 3.0mm x 14mm NL Hexalobe Screw

FDA Recall Z-2532-2026 · Posted 2026-06-23

Recalling FirmAcumed LLC
Recall StatusOpen, Classified
Product CodeHRS
Date Initiated2026-05-28
Product Quantity5,272 units

Reason for Recall

Due to manufacturing error, hexalobe screws used with wrist fixation system may break during use.

Firm Action

On May 28, 2026 Acumed issued a Urgent Medical Device Recall Notification to affected consignees via phone calls and email. Acumed asked consignees to take the following actions: 1)IMMEDIATELY stop using and/or distributing the product and complete the following actions: 2) Perform a physical count of your inventory. Identify any inventory of the batch/lot numbers listed in the table above. 3) Remove the product segregated in step 1, such that they are taken out of service (i.e., quarantined). 3) Verbally acknowledge (then respond via email) to [email protected] the recall notice. 4) You will be contacted by Field Inventory department who will directly remove the screws and conduct a count reconciliation. 5) This notice needs to be passed on to all those who need to be aware within your organization or facilities where the affected devices have been transferred.

Root Cause

Device Design

Distribution Pattern

US Nationwide distribution in the states of AR, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MS, NC, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, WA, WI.