Terminated

LEEP Return Patient Pads (10/ box) Model # 6050Pl used in electrosurgical procedures.

FDA Recall Z-2545-2019 · Posted None

Recalling FirmCooperSurgical, Inc.
Recall StatusTerminated
Product CodeGEI
Date Initiated2019-08-19
Date Terminated2020-07-14
Product Quantity4930 units

Reason for Recall

Discrepancy between the correct expiration dates on the individual pouches inside the boxes and incorrect expiration dates on the outer boxes, where the month and day were reversed.

Firm Action

Firm sent letter 8/19/2019. 1. Please discard the outer master 10- pack box containing barcode label . 2. Keep the individual seal pouches with correct expiration date. 3. Complete and return the acknowledgement and receipt form. 4. To return the product, Qurantine the affected product, complete and retun the acknowledegment and receipt form to CooperSurgical. If you have additional questions, please contact a CooperSurgical Product Surveillance representative at 203.601.5200 Ext. 3300 or email us at [email protected].

Root Cause

Under Investigation by firm

Distribution Pattern

USA: AK, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, & WV. Distribution US nationwide, Canada, Colombia, Japan, & Malaysia