Terminated

Centurion DIFFICULT IV KIT containing BD's Vacutainer Plastic Lithium Heparin Tubes (BD Vacutainer-BD Codes 367884 & 367960; Centurion code 45177 (VACUTAINER GREEN TOP) and 84815 (VACUTAINER 3ML LT. GREEN). Centurion Kit Code: IV8860

FDA Recall Z-2545-2020 · Posted None

Recall StatusTerminated
Product CodeJKA
Date Initiated2019-06-01
Date Terminated2021-06-21
Product Quantity140 units

Reason for Recall

False elevation of carboxyhemoglobin (COHb) results in blood samples collected using BD Vacutainer Plastic Lithium Heparin tubes with the IL GEM 4000 instrument.

Firm Action

Centurion Notification issued to parent company, Medline Industries, via email to Recall Team via their Corporate Headquarters on 6/18/19. Letter states reason for recall, health risk and Required Action: 1. Whether or not you have product in inventory, please complete and return the attached Accountability Record via email - [email protected] or via fax -517-546-3356. 2. Please forward the BD notice to any customer(s) that may receive these kits. 3. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program. Contact BO at 1-888-237-2762, (select Option #3 and then Option #2), between 8AM and 5PM CT Monday through Friday.

Root Cause

Device Design

Distribution Pattern

US Nationwide distribution including in the state of IL.

Official Source

View FDA Recall →