MediTrace Cadence Adult Zoll, Preconnect Defibrillation Electrode, Nonsterile Item Code:22770PC
FDA Recall Z-2546-2015 · Posted 2015-09-02
Advertisement
Recalling FirmCovidien LLC
Recall StatusTerminated
Product CodeDRO
Date Initiated2015-08-19
Date Terminated2017-05-24
Product Quantity114,000 sets
Reason for Recall
Multi-Function Defibrillation Electrodes, Pre connect-
potential for damage to the wire insulation may
result in a potentially increased risk for reduced or no patient therapy and patient and/or clinician burns
Firm Action
Medtonic/Covidien issued recall letter on August 19, 2015. Customers were notified via Federal Express and the letter instructs customers to discontinue use of the product and return all inventory to Medtronic (Covidien), 2 Ludlow Park Drive, Chicopee, MA 01022, Attention: Field Action Returns.
Contact your Medtronic representative or Customer Service at (800)-882-5878.
Root Cause
Process control
Distribution Pattern
Nationwide Foreign: Canada Australia Denmark Germany New Zealand
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.