USPACK-SUTURE KIT; Item Number: US1672
FDA Recall Z-2546-2017 · Posted None
Advertisement
Recalling FirmMedtronic
Recall StatusTerminated
Product CodeGAM
Date Initiated2017-05-03
Date Terminated2019-07-18
Product Quantity478,465 (total for all products)
Reason for Recall
Potential for an incomplete seal on the inner package, which may result in premature degradation of the suture material.
Firm Action
All consignees were notified via Federal Express or certified mail on May 3, 2017, and the letter informs customers of the potential for an incomplete seal on the inner package and the actions they should take. Customers are requested to acknowledge receipt and understanding of the Urgent Field Corrective Action Notice.
Root Cause
Process control
Distribution Pattern
Worldwide distribution. US Nationwide including PR, and countries of AU, BE, CA, CL, CN, IL, IT, JP, KR, MX, PA, SA, SG, and TR.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.