Terminated

Xpert MRSA Catalog GXMRSA-120, Rev. AJ Microbiology: The Cepheid Xpert MRSA assay performed in the GeneXpert Dx System (Xpert MRSA) is a qualitative in vitro diagnostic test designed for rapid detection of Methicillin-Resistant Staphylococcus Aureus (MRSA) from nasal swabs in patients at risk for nasal colonization

FDA Recall Z-2548-2016 · Posted None

Recalling FirmCepheid
Recall StatusTerminated
Product CodeNQX
Date Initiated2016-07-18
Date Terminated2016-09-30
Product Quantity68 kit for lot 1000037539 and 127 kit s for lot 1000029307.

Reason for Recall

Internal investigation found the certain lots of MRSA kits have the potential for performance issues at lower positive sample concentrations.

Firm Action

Recall letters dated July 18, 2016 were sent to affected customers on July 19, 2016 by Federal Express. Letters advised users of the affected product and asked that any remaining inventory be destroyed.

Root Cause

Nonconforming Material/Component

Distribution Pattern

US only: PA, DE, FL, SC, ME, CA, IL, NC, TX, WA, NY, TN, FL, OR, OK, MO, MS, GA, NM, IN, OH, MD, VA, CO, AZ, and NC.