Terminated

ImplantDirect, Dental implants intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework.

FDA Recall Z-2551-2018 · Posted None

Recall StatusTerminated
Product CodeDZE
Date Initiated2017-03-03
Date Terminated2018-08-06
Product Quantity124 units

Reason for Recall

The main vial label lists the incorrect part number, but the cap label is correct.

Firm Action

Recall notifications dated 2/13/2017 were issued to U.S. consignees via UPS overnight on 3/3/2017. Foreign consignees received letters on 3/13/2017 and 3/14/2017.

Root Cause

Process control

Distribution Pattern

Distribution was made to AZ, CA, CO, CT, FL, ID, IL, NC, NJ, NV, NY, PA, TX, UT, VA, and WA. There was no government/military distribution. Foreign distribution was made to Switzerland, Germany, Lithuania, and Spain.