Terminated

Siemens Dimension EXL Integrated Chemistry System in vitro diagnostic devices intended to measure a variety of analytes in human body fluids. All Serial #s of material #s: 10636929, 10636928, 10486890, 10472176, 10639337, and 10639338.

FDA Recall Z-2553-2014 · Posted 2014-08-26

Recall StatusTerminated
Product CodeJJE
Date Initiated2014-07-16
Date Terminated2014-12-16
Product Quantity2757

Reason for Recall

The Reagent Lid hinge may lose its effectiveness and slowly shift downward during maintenance procedures, potentially resulting in physical contact and injury to the operator.

Firm Action

Siemens notified end users of a potential safety issue with the Reagent Lid hinge on the Dimension EXL instrument via an Urgent Medical Device Correction letter, dated July 2014. The letter indicated Customer Service Engineers will inspect and adjust the tension on the reagent lid hinges during every visit and replace hinges which no longer properly adjust. Users were urged to use caution when the Reagent Lid is open and notify Siemens the Reagent Lid is shifting down.

Root Cause

Device Design

Distribution Pattern

Nationwide and internationally, including Canada and Mexico