Thuvera Laser Console. Software Version 2.2. Portable Thulium Fiber Laser.
FDA Recall Z-2554-2026 · Posted 2026-06-24
Advertisement
Recalling FirmIPG Medical Corporation
Recall StatusOpen, Classified
Product CodeGEX
Date Initiated2026-05-05
Product Quantity198 units
Reason for Recall
Software anomaly that causes a false display of error code 5018.
Firm Action
An URGENT MEDICAL DEVICE RECALL notice dated 05/05/2026 was emailed to IPG Medical's consignee notifying them of this recall correction to address Error 5018 and Error 5007/5023 displayed on monitors with software version 2.2. The notification provides instructions to consignees to work around the error messages until a software upgrade is deployed at a later date. The notification instructs consignees that their staff may correct units onsite via software upgrade or units may be returned to IPG Medical for a factory correction when the update is available. A corrected IFU will be provided to all customers. Questions can be directed via phone to 774-633-5903 or [email protected].
Root Cause
Device Design
Distribution Pattern
US Nationwide distribution in the state of MA.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.