TiBrid Stand Alone Intervertebral Body Fusion Device, TiBrid SA Angled Screwdrivers, Sizes 30 degree, 45 degree, and 60 degree angles.
FDA Recall Z-2557-2025 · Posted 2025-09-11
Advertisement
Recalling FirmOmnia Medical
Recall StatusOpen, Classified
Product CodeOVD
Date Initiated2024-07-15
Product Quantity15 units
Reason for Recall
Failure of fusion system instruments in the field.
Firm Action
Consignees were called on or around 8/1/2024 to notify them of this recall of instruments. Affected instruments were to be sent back to Omnia Medical and replaced with instruments that underwent rework in the form of design changes aimed at increasing their durability.
Root Cause
Device Design
Distribution Pattern
US Nationwide distribution in the states of AZ, CA, NJ, NV, OK & WA.