V-Twin, Model # 6002-800, UDI: 03661540600180 Product Usage: An in-vitro diagnostic device, which is an automated chemistry analyzer, micro type, intended for clinical use in conjunction with certain materials to measure a variety of analytes, including applications in clinical chemistry, monitoring drugs of abuse and in therapeutic drug monitoring.

FDA Recall Z-2558-2019 · Posted None

Recalling FirmVital Scientific N.V.
Recall StatusOpen, Classified
Product CodeJJF
Date Initiated2016-01-18
Product Quantity2

Reason for Recall

Instrument stopped working due to a software lockup, and no patient results are produced.

Firm Action

In very close cooperation with the consignees, the affected end users running on EPROM version V1.2.3 were contacted. Initially, a temporary work around (reset of the software) was communicated with the consignees. The consignees devices were installed to document all errors during runtime .In addition, a new software was developed and tested. The new software resulted in version 1.2.5 of EPROM, which was released June 23, 2016. Consignees of the V-Twin all received a field update kit with the new software.

Root Cause

Software Manufacturing/Software Deployment

Distribution Pattern

US in the state of New York.