Siemens EasyLink Informatics System V4.0/V5 Siemens, Material Number (SMN) 10444799, 10444800, 10487219 A clinical data management system that assists medical laboratory professionals with preanalytic and post-analytic functions in conjunction with multiple instruments, the laboratory information system (LIS) and Siemens StreamLAB Automation Solutions.
FDA Recall Z-2561-2014 · Posted 2014-09-01
Advertisement
Recalling FirmSiemens Healthcare Diagnostics, Inc.
Recall StatusTerminated
Product CodeJQP
Date Initiated2013-08-13
Date Terminated2014-12-02
Product Quantity1347
Reason for Recall
May not perform as intended under certain conditions, causing the release of results to the Laboratory Information System (LIS) that should have been held for manual review due to auto-verification rules or the delay/omission of result
transmission to the LIS.
Firm Action
Siemens issued an Urgent Medical Device Correction (UMDC) letter on August 13, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to review this letter in its entirety and ensure that all operators understand the information presented. The issues described are specific to the features listed in the letter.. If the laboratory is not using a particular feature, then they are not affected by the issues described for that feature. For questions contact your Siemens Customer Care Center or your local Siemens technical support
representative.
Root Cause
Software design
Distribution Pattern
US Nationwide Distribution