1) TVS LPA, Sterile EO, Rx Only, Catalog #: a) KIT-071-01; b) KIT-071-03; c) KIT-071-10; d) KIT-071-11; 2) Enhance Transcarotid/Peripheral Access Kit, Sterile EO, Rx Only, Catalog #: a) KIT-075-02; b) KIT-075-03

FDA Recall Z-2561-2018 · Posted None

Recall StatusTerminated
Product CodeDRE
Date Initiated2018-05-02
Date Terminated2020-07-20
Product Quantity8350

Reason for Recall

The products listed might contain unsafe levels of bacterial endotoxins (Pyrogens) that were introduced during a manufacturing step. Bacterial Endotoxins also called pyrogenic bacteria can activate the inflammatory process and produce fever, chills, and hypotension in a patient.

Firm Action

On approximately 05/01/2018 recall letters and customer response forms were mailed to all consignees.

Root Cause

Other

Distribution Pattern

U.S.: MN, PA, VA, CA, UT, FL, NY, GA, LA, MO, TX, MS, OH, WA, MA, IL, KY, NC, and Puerto Rico; OUS: Canada, Australia, France, Germany, Ireland, Italy, Netherlands, Poland, Slovakia, Switzerland, Belgium, United Kingdom, Korea