LYMOL Medical Elite X Class Rigid Bronchoscope System, Part No. BD2410, BB2701, BB2702, BB2401, BB2402, BA2700, BT2000, BT2101, BT2103, BT2105, BT2106, BT2210, BT2201, BT2203, BT2205, BR2200, BP2200, BI2200, BR2000, BP2000, BI2999, BR2103, BP2103, BI2103, BR2101, BP2101, BI2101, BI2001, BDL2420
FDA Recall Z-2562-2021 · Posted None
Advertisement
Recalling FirmLymol Medical Corporation
Recall StatusTerminated
Product CodeEOQ
Date Initiated2021-07-29
Date Terminated2022-10-04
Product Quantity309
Reason for Recall
Reprocessing instructions were revised November 15, 2020, to reflect new processes and equipment used at consignee user facilities. New instructions were not sent to all consignees of the devices at the time of the revision.
Firm Action
On August 9, 2021, the firm began notifying all consignees via email that the reprocessing instructions have changed. Included with each notification is a copy of the Urgent Notice, the revised IFU, and a response form.
Customers were advised that the sterilization procedures were updated on November 15, 2020. Customers should discard and replace the obsolete procedures with the revised procedures included with the firm's mailing.
The revised instructions have been uploaded to the LYMOL website at https://www.lymolmedical.com/instructions.cfm
Customers with questions may contact the firm at [email protected].
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide distribution.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.