LYMOL Medical Elite X Class Rigid Bronchoscope System, Part No. BD2410, BB2701, BB2702, BB2401, BB2402, BA2700, BT2000, BT2101, BT2103, BT2105, BT2106, BT2210, BT2201, BT2203, BT2205, BR2200, BP2200, BI2200, BR2000, BP2000, BI2999, BR2103, BP2103, BI2103, BR2101, BP2101, BI2101, BI2001, BDL2420

FDA Recall Z-2562-2021 · Posted None

Recall StatusTerminated
Product CodeEOQ
Date Initiated2021-07-29
Date Terminated2022-10-04
Product Quantity309

Reason for Recall

Reprocessing instructions were revised November 15, 2020, to reflect new processes and equipment used at consignee user facilities. New instructions were not sent to all consignees of the devices at the time of the revision.

Firm Action

On August 9, 2021, the firm began notifying all consignees via email that the reprocessing instructions have changed. Included with each notification is a copy of the Urgent Notice, the revised IFU, and a response form. Customers were advised that the sterilization procedures were updated on November 15, 2020. Customers should discard and replace the obsolete procedures with the revised procedures included with the firm's mailing. The revised instructions have been uploaded to the LYMOL website at https://www.lymolmedical.com/instructions.cfm Customers with questions may contact the firm at [email protected].

Root Cause

Under Investigation by firm

Distribution Pattern

US Nationwide distribution.

Official Source

View FDA Recall →