Valleylab Laparoscopic Curved Spatula Electrode 45CM Item Code: E277145 The Valleylab Laparoscopic Handset and Electrodes are indicated for use in laparoscopic and thoracoscopic surgical procedures where the delivery of electrosurgical current, irrigation and suction via a single handset is desirable
FDA Recall Z-2563-2016 · Posted None
Recalling FirmMedtronic
Recall StatusTerminated
Product CodeGEI
Date Initiated2016-07-08
Date Terminated2018-08-08
Product Quantity6381
Reason for Recall
Potential for compromise of the package resulting in a breach of the sterile barrier
Firm Action
Medtronic/Covidien sent an Urgent Medical Device Recall letter dated July 8, 2016, to all affected customers via Federal Express. The letter identified the product, the problem, and the action to be taken by the customer. The letter instructs customers to discontinue use of the product and return all inventory to Medtronic (Covidien), 195 McDermott Road, North Haven, CT 06473, Attention: Field Returns Department. Customers with questions were instructed to contact their Medtronic representative or Customer Service at 800-882-5878.
Root Cause
Process control
Distribution Pattern
Worldwide Distribution - US (nationwide) and Internationally to Australia, Canada, Japan, Korea, Hong Kong, Thailand,
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.