Medsolutions4U Multi-Function Electrodes to be used with PhysioControl Lifepak Defibrillators, models 9, 10C, 11, 12, 15, 20, 500 and are indicated for Defibrillation, Cardioversion, Pacing, and Monitoring Part Number: 2033
FDA Recall Z-2565-2014 · Posted 2014-09-02
Advertisement
Recalling FirmBio-Detek, Inc.
Recall StatusTerminated
Product CodeMKJ
Date Initiated2013-03-12
Date Terminated2015-06-17
Product Quantity30 cases (300 electordes)
Reason for Recall
Medsoultions4U Multi-Function Lifepak Electrodes, manufactured with an incorrect wire connector which could not interface with the defibrillators
Firm Action
Bio-Detek issued an Urgent Field Safety Notice to Medsolutions4 U on 3/12/13, informing them of the problem and actions that should be taken to remove the affected products from service.
Root Cause
Nonconforming Material/Component
Distribution Pattern
Worldwide Distribution in Germany and Slovenia.