Recall StatusTerminated
Date Initiated2018-06-08
Date Terminated2018-11-05
Product Quantity1 kit (US), 602 kits (OUS)
Reason for Recall
The device is not suitable for the control of the Troponin T assay due to unacceptable variation between vials. The deviation in precision observed is up to 28%. IQC that is reported as out of range could lead to a delay in reporting Troponin T results. If the failure occurs the operator may observe the quality control result for Troponin T falling outside range and/or increased imprecision for Quality Control monitoring.
Firm Action
On June 8, 2018, the firm issued an Urgent Medical Device Recall letter to affected consignees. The letter stated that Randox has confirmed Liquid Cardiac Control CQ5053 Lot: 4245CK is not suitable for the control of
the Troponin T assay due to unacceptable variation between vials. The deviation in precision observed is up to 28%.
Customers were instructed to do the following:
" Review results generated with the affected batches in line with the clinical profile of the patient.
" Discuss the contents of this notice with your Medical Director.
" Update kits with corrected IFU's excluding Troponin T values.
" Complete and return the vigilance response section of this form to
[email protected] within five working days.)
In addition, distributors were instructed to send a copy of the letter to all affected customers and to those who need to be aware within their organization.
Root Cause
Under Investigation by firm
Distribution Pattern
Worldwide Distribution - US in the state of WV. Foreign distribution to Australia, Chile, China, Germany, Hungary, Ireland, Italy, Moldova, Netherlands, Peru, Poland, Saudi Arabia, Slovakia, Spain, Sultanate of Oman, UAE, UK, and Vietnam
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source