Terminated

Vysis CLL FISH Probe Kit with the following components: Vysis LSI p53 SpectrumOrange/ATM, SpectrumGreen and LSI D13S319, SpectrumOrange/ 13q34, SpectrumAqua/CEP 12 SpectrumGreen Probes

FDA Recall Z-2576-2021 · Posted None

Recall StatusTerminated
Product CodeOVQ
Date Initiated2021-08-04
Date Terminated2023-05-31
Product Quantity186 units

Reason for Recall

Potential for Vysis CLL FISH Probe Kits not detecting 13q deletions in known positive patient samples.

Firm Action

On about 08/04/2021, Abbott Molecular notified consignees via FedEx with letter titled "Urgent Field Safety Notice Molecular Diagnostics at Abbott Product: Vysis CLL FISH Probe." The letter instructed customers to discontinue use of the affected lots, discard any remaining kits on hand, complete and return the customer reply form, and notify impacted customers if the affected lots have been further distributed. Additional instructions included to review the recall notification information with your Medical Director or physicians as appropriate and retain the communication for future reference. Review patient results generated with the impacted lots and determine if retesting is required taking patient medical history and previous treatment into consideration.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide distribution - US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, KY, MI, MN, MA, NC, NJ, NM, NY, OH, PA, TN, TX, UT, VA, VT, WA, WI and the countries of Australia, Belgium, Brazil, Canada, China, Costa Rica, Czech Republic, Estonia, Germany, Hong Kong, Israel, Italy, Kazakhstan, Netherlands, Philippines, Poland, Saudi Arabia, Slovenia, South Africa, South Korea, Spain, United Kingdom, Uruguay.

Official Source

View FDA Recall →