Philips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following system descriptions: TELE Monitor 5500, 1.4 Ghz, ECG only, ex; Telemetry Monitor 5500 1.4 GHz FAST EX.
FDA Recall Z-2579-2026 · Posted 2026-06-24
Advertisement
Recalling FirmPhilips North America Llc
Recall StatusOpen, Classified
Product CodeMHX
Date Initiated2026-06-05
Product Quantity909 units
Reason for Recall
Potential for device reset to default "NEW_DEVICE" state, which may lead to a to loss of device configuration and the equipment label, resulting in a failure to reconnect to the PIC iX central monitoring system.
Firm Action
Philips Medical notified consignees on about 06/05/2026 via "URGENT Medical Device Correction" letter. Consignees were notified of the issue and potential hazard and instructed to continue the use of the device but avoid touching the display while the device is booting up and provide this notification to all personnel who need to be aware within each organization or to any organization where affected product(s) have been potentially transferred.
Philips will release new software which will correct this problem. A Philips representative will contact you to arrange for an upgrade to the new software.
Root Cause
Software design
Distribution Pattern
US Nationwide distribution in the states of AZ, CA, CO, IN, MA, MI, MN, PA, SD, TN.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.