Terminated

PERSONA" The Personalized Knee System UC Tibial Articular Surface Provisional Left Size 3-7 CD Top; Left Size 4-11 EF Top; Left Size 7-12 GH Top; Left Size 9-12 J Top; Right Size 3-7 CD Top; Right Size 4-11 EF Top; Right Size 7-12 GH Top; and Right Size 9-12 J Top.

FDA Recall Z-2580-2014 · Posted 2014-09-04

Recalling FirmZimmer, Inc.
Recall StatusTerminated
Product CodeJWH
Date Initiated2014-08-07
Date Terminated2015-01-27
Product Quantity77,737 total units

Reason for Recall

Reports of breakage of the Persona Knee Trial Articular Surface Provisionals (TASPs). The Persona primary knee surgical technique has been updated. This field action correction is to ensure that all distributors and users of this instrument are aware of the enhanced surgical techniques.

Firm Action

On 8/7/2014, Zimmer notified all distributors via electronic mail. Distributors that have received affected inventory are also notified via courier. Hospital risk managers and surgeons are also notified via courier. Hospital risk managers and surgeons were provided with a notification identifying the issue and their responsibilities. These responsibilities include ensuring the affected personnel are aware of the contents of the notification. Distributors were sent a notification identifying the issue and to provide a copy of the notification sent to surgeons and risk managers. The distributors will also be instructed as needed to assist with identifying a comprehensive list of surgeons.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution: US (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY; and countries of: AUSTRALIA, INDIA, JAPAN, KOREA, SINGAPORE, AUSTRIA, BELGIUM, FRANCE, GERMANY, ISRAEL, ITALY, LUXEMBOURG, NETHERLANDS, SAUDI ARABIA, SOUTH AFRICA, SPAIN, SWITZERLAND, UNITED KINGDOM, and UNITED ARAB EMIRATES.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →