Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

i-STAT CG4+ cartridge (white). List Number: 03P85-25.

FDA Recall Z-2582-2025 · Posted 2025-09-17

Recall StatusOpen, Classified
Product CodeCHL
Date Initiated2025-08-21
Product Quantity9,067,050 units

Reason for Recall

Lack of a 510(k) premarket clearance for the i-STAT CG4+ cartridge to reflect updated sample type for lactate and measurement ranges for pH and PO2.

Firm Action

Abbott notified consignees via email and letter on about 08/21/2025. Consignees were instructed to download, review and implement the updated i-STAT CG4+ cartridge Instructions for Use (IFU), discontinue use of the i-STAT CG4+ cartridges for testing of capillary whole blood samples for lactate and instead use venous or arterial whole blood or an alternate method for testing capillary whole blood for lactate, and update the pH and PO2 reportable ranges on the i-STAT 1 analyzers running i-STAT CG4+ cartridges. This feature is not customizable through the i-STAT 1 handheld keypad, please see the i-STAT DE Quick Reference Guide - Customizing Reportable Ranges (ART 770547-00 Revision A). To use the Custom Reportable Range feature, you must have the i-STAT 1 handheld and i-STAT DE version 2.3 or higher. Changes to the reportable range may require a change to the Laboratory Information System (LIS) interface. Additionally, consignees were requested to confirm receipt and understanding of this communication by responding to the business reply card included with the letter. If consignees have forwarded any i-STAT CG4+ cartridges to another facility, they were requested to notify them.

Root Cause

No Marketing Application

Distribution Pattern

US Nationwide distribution.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑