Open, Classified

Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. BARIATRIC DYNJ905153J DYNJ905153K DYNJ905153L GYNE LAPAROTOMY PACK-LF DYNJ49593A LAPAROSCOPY DYNJ905157G DYNJ905157I DYNJ905157J LAPAROSCOPY PACK DYNJ56930B MAJOR PLUS PACK DYNJ905160J DYNJ905160K DYNJ905160M MINIMALLY INVASIVE PACK-LF DYNJ0843063K PERCUTANEOUS TRAY DYNJ80196 ROBOTIC PACK DYNJ908706 DYNJ908706B

FDA Recall Z-2583-2026 · Posted 2026-06-25

Recall StatusOpen, Classified
Product CodeFDE
Date Initiated2026-02-27
Product Quantity113,843 kits

Reason for Recall

Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.

Firm Action

This entry catalogues SKUs communicated through the same letter documented in RES 98601 which were inadvertently missed in the initial report to the FDA. Firm began notifying customers on February 27, 2026 via letters titled "URGENT MEDICAL DEVICE RECALL". Customers were instructed to immediately check stock for affected product and quarantine any affected units. Upon receipt of the customers' response form reporting the quantity of affected product on hand, the firm will provide over-labels to place on affected inventory. Labels contain instructions for staff to remove and discard the affected component prior to using the kit.

Root Cause

No Marketing Application

Distribution Pattern

US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI, WY and the country of Barbados.

Official Source

View FDA Recall →