CONCEPT GRAFIX TENDON STRIPPER. REF TS8850. 7.0MM, QTY 1. Rx Only, Non-Sterile, Made In USA.

FDA Recall Z-2585-2010 · Posted 2010-09-28

Recall StatusTerminated
Product CodeFZW
Date Initiated2010-01-29
Date Terminated2011-01-12
Product Quantity149

Reason for Recall

There is a possibility that the device may exhibit Hydrogen Embrittlement in the material and are subject to potential breakage.

Firm Action

Recall Notification Letters were sent to customers on January 29, 2010. Customers were requested to complete and return a Customer Response Form indicating the number of products available for return and number distributed. Each distributor was asked to notifiy their customers if the product was further distributed.

Root Cause

Process control

Distribution Pattern

Distributed throughout U.S. and Internationally to Arab Emirates, Belgium, Canada, China, Columbia, Cyprus, Germany, Denmark, Egypt, Spain, France, United Kingdom, Greece, Hong Kong, India, Italy, Japan, Republic of Korea, Lithuania, Morocco, Malaysia, Poland, Portugal, Saudi Arabia, El Salvador, Thailand, Taiwan and Viet Nam. Domestic Distribution to CA, FL, GA, IL, IN, KS, MI, MT, NY, OH, PA, TX, and WA.