OPTICAL BIOMETER AL-Scan: Software Version: 1.09 and earlier. Ophthalmic: The OPTICAL BIOMETER AL-Scan is a measuring device that can measure as a single unit the values necessary to calculate the power of an IOL for cataract surgery such as axial length, corneal curvature radius, and anterior chamber depth. It measures these necessary values successively through a non-contact optical measurement method.

FDA Recall Z-2587-2016 · Posted None

Recalling FirmNidek Inc
Recall StatusTerminated
Product CodeHJO
Date Initiated2016-07-01
Date Terminated2016-12-22
Product Quantity17

Reason for Recall

Nidek Inc. received information from our manufacturer NIDEK CO. Japan that the shown axial lengths were much shorter at AL-Scan measurement.

Firm Action

Dear Doctor/Recall letters were sent to affected customers on July 7, 2016. The letters describe the issue and provide instruction should the issue occur. The letters inform users that Nidek will reach out in the next few weeks to schedule installation of new software that will solve the problem. Questions or concerns may be directed to: Customer Service 1.800.722.0219; Monday- Friday 8 am to 5pm PST E-mail: [email protected].

Root Cause

Software in the Use Environment

Distribution Pattern

US Nationwide Distribution in the states of: NY, PA. GA. AZ, VA, DE, ME,. MO, TN, and NM.