Philips Allura Xper, Allura Centron, Allura CV, Azurion, UNIQ Clarity, and MultiDiagnost-Eleva systems

FDA Recall Z-2587-2023 · Posted 2023-09-14

Recall StatusOpen, Classified
Product CodeOWB
Date Initiated2023-08-15
Product Quantity19,115 systems

Reason for Recall

A damaged foot switch may cause instances where no or only intermittent x-ray radiation initiation is possible.

Firm Action

US customers were notified via certified mail via USPS. OUS customers were notified through the Philips Markets Organizations. Philips is issuing two Urgent Medical Device Correction letters, one for Allura and Azurion customers and another one for MultiDiagnost Eleva Customers. Additionally, an IFU addendum will be provided to customers that includes instructions on how to handle the footswitch cable. For Allura and Azurion customers an updated preventative maintenance manual update is also being provided. Philips will contact customers to schedule a visit to inspect the foot switch and to provide an IFU Addendum.

Root Cause

Device Design

Distribution Pattern

Worldwide distribution