Terminated

PROFEMUR(R) MODULAR NECK, HIP FEMORAL NECK MicroPort total hip systems intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients.

FDA Recall Z-2588-2016 · Posted 2016-08-18

Recall StatusTerminated
Product CodeLWJ
Date Initiated2016-06-30
Date Terminated2017-01-21
Product Quantity14 units

Reason for Recall

Profemur(R) neck, Item Number PHA01222 (Short AR VV2 Ti Neck) was labeled as Item Number PHA01212 (Short AR VV1 Ti Neck).

Firm Action

The firm initiated their recall by email on 06/30/2016.

Root Cause

Error in labeling

Distribution Pattern

Belgium, Italy, Spain, and Poland