Terminated

GALT VALVED TEARAWAY KIT GALT VTI w/ Sideport Stainless Steel Dbl Flex. J-St RX ONLY UNITS 1 STERILE EO Product Usage: Used in percutaneous procedures to introduce catheters and other intravascular devices into the peripheral vasculature.

FDA Recall Z-2589-2014 · Posted 2014-09-04

Recall StatusTerminated
Product CodeDYB
Date Initiated2014-07-25
Date Terminated2016-02-26
Product Quantity11,155 units in total

Reason for Recall

The adhesive used in the product could dislodge when objects are inserted through the valve which could pose an embolic risk to the patient.

Firm Action

Galt Medical sent an Urgent Medical Device Recall letters to all affected customers. The letter identified the affected product, reason for recall and actions to be taken. Customers were asked to identify and segregate the recalled lot (s) that is in their possession, complete and return the enclosed Recall Reply form, and return the recalled product to Galt Medical Corp. For questions contact your local sales representative or Recall Coordinator at 214-778-1306.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution - US (nationwide) Canada, Germany, the Netherlands, Ireland, Switzerland, France, Italy, and Greece.