Terminated

iTRACK250A Canaloplasty Microcatheter Kit; Catalogue number: iT-250A; Rx only, sterile. Indicated for fluid infusion and aspiration during surgery.

FDA Recall Z-2590-2014 · Posted 2014-09-04

Recalling FirmEllex iScience, Inc.
Recall StatusTerminated
Product CodeHMX
Date Initiated2014-08-06
Date Terminated2014-10-28
Product Quantity148 devices

Reason for Recall

One lot may not have been properly sealed, resulting in a non-sterile device.

Firm Action

Urgent Medical Device Recall letters were sent to all customers on August 6, 2014 by trackable mail. Letters advised on the potential lack of sterility of iTRACK250A, Canaloplasty Microcatheter Kit, lot 1404-03 due to improperly sealed pouches. Lack of sterility may pose the risk of infection to the patient. Customers are identify and segregate any affected product in their inventory. The affected product should be returned following the instructions provided. Replacement product will be provided at no charge. Questions should be directed to Customer Care at 1-888-846-4724, M-F 8 am to 4:30 Pm Eastern Time.

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide Distribution -- US (nationwide) and countries of: Hong Kong and Switzerland.