Merge PACS software. Merge PACS (Picture Archiving Communication System) is designed and marketed for soft copy reading, communication, and storage of studies produced by digital modalities, including digital mammography
FDA Recall Z-2591-2017 · Posted None
Advertisement
Recalling FirmMerge Healthcare, Inc.
Recall StatusTerminated
Product CodeLLZ
Date Initiated2017-05-08
Date Terminated2019-08-08
Product Quantity47 sites have the potentially affected versions
Reason for Recall
When cut lines are shown on an image, the lines may go from being vertical and then shift to horizontal causing confusion to the physician. This may cause a misdiagnosis.
Firm Action
Merge sent an Urgent Medical Device Recall letter dated May 5, 2017, to all affected customers. Customers were instructed to be aware of the following workaround: Do not have the simple close study red x icon available in the reading physicians toolbars. This will force them to set status on close which also presents the warning Not all images have been reviewed, do you want to set status anyway and close study or continue reviewing unviewed images?
YOUR RESPONSE TO THIS NOTIFICATION IS REQUIRED
By replying to this notification, you will be taking the necessary steps to ensure that the issue listed in this document will be resolved.
Please reply using the enclosed form and the return addressed envelope.
Customers with questions were instructed to send an email to [email protected].
For questions regarding this recall call 877-741-6369.
Root Cause
Software design (manufacturing process)
Distribution Pattern
Nationwide Distribution
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.