Open, Classified

Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. ANGIO DRAPE PACK (ADLUG)642-LF DYNJ47665D ANGIO PACK DYNJ30565D ANGIOGRAPHY PACK-LF DYNJ0854485V ARTERIOGRAM DYNJ28082D DYNJ28082F CARDIAC CATH DYNJ67349 CARDIAC CATH PACK DYNJ57874C CATH I PLUS-LF DYNJSCL100 CATH III PLUS-LF DYNJSCL300 CATH LAB ANGIOGRAPHY PACK DYNJ50779F CATH LAB PACK DYNJ41088B DYNJ48085 CATH LAB TRAY DYNJ37389N DYNJ37389O CATH PACK DYNJ62595A DYNJ69997 EP DRAPE PACK-LF DYNJ47645B HOSPITAL REGIONAL DE CONCEPCIO DYNJ67918A INTERVENTIONAL PACK DYNJ56666C INTERVENTIONAL RAD PACK DYNJ44100L INTERVENTIONAL RADIOLOGY PACK DYNJ43664 INTERVENTIONAL RADIOLOGY PK-LF DYNJ0516014AF OR HYBRID-MRMC DYNJ907758A VMA TRAY DYNJ62376A

FDA Recall Z-2592-2026 · Posted 2026-06-25

Recall StatusOpen, Classified
Product CodeOES
Date Initiated2026-02-27
Product Quantity113, 843 kits

Reason for Recall

Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.

Firm Action

This entry catalogues SKUs communicated through the same letter documented in RES 98601 which were inadvertently missed in the initial report to the FDA. Firm began notifying customers on February 27, 2026 via letters titled "URGENT MEDICAL DEVICE RECALL". Customers were instructed to immediately check stock for affected product and quarantine any affected units. Upon receipt of the customers' response form reporting the quantity of affected product on hand, the firm will provide over-labels to place on affected inventory. Labels contain instructions for staff to remove and discard the affected component prior to using the kit.

Root Cause

No Marketing Application

Distribution Pattern

US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI, WY and the country of Barbados.