Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Convenience kits containing select SKUs of 10mL Polycarbonate Colored Syringes. ABDOMINAL VASCULAR-LF DYNJ905291B DYNJ905291C ARTERIOGRAM DYNJ28082G CVICU HEART BAG DYKM2130A ENDOVASCULAR PACK DYNJ0678934J EP PACK DYNJ23456J GENERAL ENDO PACK-LF DYNJ0842873K HYBRID PACK DYNJ65925B DYNJ65925C NON VASCULAR PACK DYNJ68037 PACEMAKER DYNJ901681D PACEMAKER PACK DYNJ51853C DYNJ51853D DYNJ60606A PACER PACK DYNJ40952A RADIOLOGY VASCULAR ACCESS TRAY DYNJVB1303A SM- INTERVENTIONAL DRAPE PK-LF DYNJ0780143K VASCULAR ACCESS PACK-LF DYNJ0101291D VASCULAR DSC DYNJ69172 DYNJ69172B VASCULAR HYBRID DYNJ57543F DYNJ57543I DYNJ57543J VASCULAR PACK DYNJ61702A

FDA Recall Z-2593-2026 · Posted 2026-06-25

Recall StatusOpen, Classified
Product CodeOEZ
Date Initiated2026-02-27
Product Quantity113, 843 kits

Reason for Recall

Firm identified unapproved design changes to the products outside of the 510(k) cleared by the FDA. Products not initially reported in RES 98601.

Firm Action

This entry catalogues SKUs communicated through the same letter documented in RES 98601 which were inadvertently missed in the initial report to the FDA. Firm began notifying customers on February 27, 2026 via letters titled "URGENT MEDICAL DEVICE RECALL". Customers were instructed to immediately check stock for affected product and quarantine any affected units. Upon receipt of the customers' response form reporting the quantity of affected product on hand, the firm will provide over-labels to place on affected inventory. Labels contain instructions for staff to remove and discard the affected component prior to using the kit.

Root Cause

No Marketing Application

Distribution Pattern

US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, TN, TX, VA, WI, WY and the country of Barbados.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑