Prime Care Transcend Mattress, Prime Care Transcend Perimeter Mattress, Prime Care Transcend Advanced Mattress, Prime Care Transcend Advanced Perimeter Mattress, Prime Plus Bariatric Mattress and Prime Plus Bariatric Perimeter Mattress, individual units packaged seperately. Model Numbers: TCM3576, TCM3580, TCM3580P, TCM3584, TCM4280, TCM4280P, TCM4284, TACM3576, TACM3580, TACM3580P, TACM3584, PPBM396, PPBM426, PPBM426P, PPBM486, PPBM546, PPBM397, PPBM427, and PPBM487. Intended to protect inner foam layers of mattress from any fluids that may be on the mattress.

FDA Recall Z-2596-2014 · Posted 2014-09-09

Recalling FirmPrimus Medical LLC
Recall StatusTerminated
Product CodeFMW
Date Initiated2014-08-11
Date Terminated2017-07-19
Product Quantity1,557 units

Reason for Recall

The firm found that the top cover of the mattresses were delaminating.

Firm Action

The firm sent out Urgent: Voluntary Field Correction/Removal letters, dated August 11, 2014, to all consignees with instructions for correcting the issue. Customers were to immediately examine their inventory and quarantine any affected product. Customers are to complete and return the response form provided. Questions can be directed to 1-877-638-2776.

Root Cause

Component design/selection

Distribution Pattern

Worldwide Distribution -- USA, including the states of TX, OH, PA, KS, FL, NV, IL, WA, MN, NJ, NC, UT, NY, CA, MD, CO. NE VA, and AZ; and, the country of Canada.

Official Source

View FDA Recall →