Terminated

BD FACS Universal Loader (Optional) to the BD FACSLyric Clinical System, Model 662879 (Universal Loader), Rx Only. The firm name on the label is Becton, Dickinson and Company, BD Biosciences, San Jose, CA. The Loader offers various settings to resuspend and mix samples. It can draw from 12 x 75-mm tubes in 30- and 40-tube racks. A barcode reader verifies the ID on tube racks and individual tubes in 30- tube racks. A built-in imaging system provides safety checks, such as verifying the correct rack type and tube layout, and ensures the tubes were loaded correctly.

FDA Recall Z-2598-2018 · Posted 2018-07-13

Recall StatusTerminated
Product CodeOYE
Date Initiated2018-05-16
Date Terminated2018-08-13
Product Quantity8 devices

Reason for Recall

The solenoid plunger tip length is too short resulting in insufficient engagement to the Universal Loader enclosure to lock during use.

Firm Action

The firm, BD Life Sciences, produced the "URGENT PRODUCT RECALL-CORRECTION" letter dated 4/6/2018 to its customers during an on-site visit. by the field service engineer made beginning 5/16/2018 to correct the issue. The visits were completed 6/15/2018. If you require further assistance, contact BD Customer/Technical Support BD Customer Support at 877.232.8995 (prompt 2) for instruments in the United States. For customers outside the US, contact your local BD representative or distributor.

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Distribution was made to FL, KS, KY, MA, NJ, and SD.