Terminated

MicroScan Pos Combo Panel Type 43, Catalog number B1017-217for use in determining antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic gram-positive cocci and Listeria monocytogenes.

FDA Recall Z-2599-2018 · Posted None

Recalling FirmBeckman Coulter Inc.
Recall StatusTerminated
Product CodeLTT
Date Initiated2018-05-04
Date Terminated2020-08-07
Product Quantity686 boxes

Reason for Recall

The drug dilution sequence for Moxifloxacin printed on the panel is incorrect. As a result, incorrect MIC results for Mxf could be reported.

Firm Action

A recall notification letter was sent the week of 05/01/2018 via email, mail, and phone if needed.

Root Cause

Labeling design

Distribution Pattern

NJ, TX, IA, PA, FL, NE, OH, NY, CA, MO, OK, LA