Prism Laser 7642 Laser System
FDA Recall Z-2600-2011 · Posted 2011-08-29
Advertisement
Recalling FirmAmerican Stress Technologies, Inc.
Recall StatusTerminated
Product CodeREO
Date Initiated2011-04-27
Date Terminated2016-10-13
Product Quantity1
Reason for Recall
It was discovered The Prism 7642 was not certified at the time of delivery. The product lacks the required labeling and the product was sold and delivered without the submission of the Initial Product Report. The product lacks User In formation
Firm Action
American Stress Technologies sent a customer notification letter dated April 1, 2011, informing them of the action they have taken to correct this issue.
(1) ASTI conducted laser safety tests and checks on May 3,2011 to verify the Prism was compliant with FDA regulations. This was confirmed in a June 17, 2011, electronic mail message.
(2) ASTI placed the required labels described in the initial product report on the Prism on May 3, 2011.
(3) ASTI delivered updated user manuals to UNO on May 3, 2011, that describe potential laser hazards and the required labeling in the same way as described in the initial product report.
(4) ASTI completed the above corrective actions at no cost to UNO as confirmed in a June 20, 2011, electronic mail message.
If you have any further questions please call (724) 410-1030
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
Nationwide Distribution
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.