Siemens RAPIDPoint 500 v2.2.2A Software Upgrade Kit; SMN 11066719, Software Version 2.2A The RAPIDPoint 500 system is designed for professional use in a point-of care or laboratory environment. This system tests blood gases, electrolytes, metabolites, total hemoglobin, and hemoglobin derivatives in arterial, venous, and capillary whole blood samples. The following parameters are tested: pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, F02Hb, FCOHb, FMetHb, FHHb, nBili
FDA Recall Z-2601-2016 · Posted 2016-08-19
Advertisement
Recalling FirmSiemens Healthcare Diagnostics Inc
Recall StatusTerminated
Product CodeKHP
Date Initiated2016-07-05
Date Terminated2017-08-02
Product Quantity76
Reason for Recall
Some v2.2.2 upgrade kits include a dialysate mode which not cleared/approved for shipment in the United States
Firm Action
Siemens sent a Customer Notification dated July 2016 to all affected domestic (US) customers via Federal Express on July 5, 2016. Customers were asked to take the following actions: Please do not use the dialysate mode. If you have enabled this feature, please disable it by following these instructions:
1. At the System screen, select Setup. 2. Select Secured Options > Analysis Options. 3. Unselect Dialysate. 4. Select the Continue button 3 times.
Please review this letter with your Medical Director. Complete and return the Field Correction Effectiveness Check attached to this letter within 7 days.
Customers with questions were instructed to contact their Siemens Customer Care Center or their local Siemens Technical Support representative.
For questions regarding this recall call 781-269-3000.
Root Cause
Under Investigation by firm
Distribution Pattern
Nationwide Distribution to AL, CA, MA, NJ, OH, TX, WA, PA