Terminated

Modera Modular Pedicle Screw System Shank Insertion Instrument, 3-107-07

FDA Recall Z-2604-2018 · Posted None

Recall StatusTerminated
Product CodeMNI
Date Initiated2017-02-13
Date Terminated2018-11-07
Product Quantity6 pieces

Reason for Recall

This is a retrospective recall of an instrument used in spine fusion procedures with Modera Modular Pedicle Screw system implants. The firm received a customer complaint in which excessive forces were applied to the shank insertion instrument and the laser weld failed.

Firm Action

On February 13, 2017, the firm emailed its distributors to notify them that their Rev. A devices would be replaced with Rev. B devices, and asked the distributors to mail back the Rev. A devices.

Root Cause

Device Design

Distribution Pattern

2 distributors in CO and TX.

Official Source

View FDA Recall →