DJO Surgical: Reverse Shoulder System (RSP) Semi Humeral Socket Insert, 32mm, REF: 508-01-032; Baseplate, Glenoid HA-Coat, RSP, 6.5mm X 30mm, REF: 508-32-104; Reverse Shoulder System (RSP) Glenoid, Neutral Head W/ Retaining Screw, 36mm, REF: 508-36-101

FDA Recall Z-2606-2023 · Posted 2023-09-19

Recalling FirmEncore Medical, LP
Recall StatusOpen, Classified
Product CodePHX
Date Initiated2023-08-04
Product Quantity64

Reason for Recall

Orthopedic device components were not subjected to final sterilization before shipment, which if implanted may lead to varying degrees of infection, which may require treatment, prolonged monitoring, and may lead to revision surgery

Firm Action

On 8/4/23, recall notices were mailed to surgeons and distributors. They were reminded that starting on 7/3/23 they were contacted by the firm, and asked to quarantine affected devices, and that the affected devices had since been retrieved by the firm's sales representatives. Surgeons were asked to take the following actions: 1) We ask that you reach out to your patient[s] who were implanted with affected components as identified above and consider whether it may be appropriate for your patient to be monitored more closely to watch for signs and symptoms of acute or chronic infection. 2) We have also provided a patient letter for you to provide to your patient[s] who was [were] implanted with an affected component[s] should you prefer a written notification to accompany your verbal contact. 3) Please make the hospital where you implanted the components aware of this action, as appropriate by providing a copy of this communication or verbally notifying the hospital. 4) Please review and complete the enclosed Voluntary Recall Response Form. If you have any questions phone the firm at 1(800)456-8696 between the hours of 7:30AM-6:00PM CST, or e-mail [email protected] In addition, distributors were asked to do the following: Provide the notification to any additional organizations or members within your facility or network where devices subject to this voluntary recall may have been transferred.

Root Cause

Nonconforming Material/Component

Distribution Pattern

US: AL, AZ, CA, CO, FL, IL, IN, KS, KY, LA, MA, MI, MN, MO, NC, NJ, NY, OH, PA, PR, RI, SC, TX, UT, VA, WA. OUS: Sweden, Australia, France, UK, Colombia, Germany