Medline Convenience kits containing recalled Swan-Ganz Catheters CARDIAC MAJOR DYNJ901328G HEART A DYNJ903465K OPEN HEART DYNJ909090G SILVER CROSS OPEN HEART DYNJ906102M VPH OPEN HEART DYNJ902664M DYNJ902664N
FDA Recall Z-2607-2026 · Posted 2026-06-26
Advertisement
Recalling FirmMedline Industries, LP
Recall StatusOpen, Classified
Product CodeOEZ
Date Initiated2026-05-28
Product Quantity9,272 kits
Reason for Recall
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Firm Action
On May 28, 2026, the firm notified customers via letters titled "URGENT MEDICAL DEVICE CORRECTION."
Customers were referred to BD's recall notice, which provides advice for users and instructions to mitigate hub breakage issues. Product return is not required.
If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Root Cause
Process change control
Distribution Pattern
US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.