Medline Convenience kits containing recalled Swan-Ganz Catheters CENTRAL AND ARTERIAL LINE TOTE DYNJ8407
FDA Recall Z-2608-2026 · Posted 2026-06-26
Advertisement
Recalling FirmMedline Industries, LP
Recall StatusOpen, Classified
Product CodeOFF
Date Initiated2026-05-28
Product Quantity9,272 kits
Reason for Recall
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Firm Action
On May 28, 2026, the firm notified customers via letters titled "URGENT MEDICAL DEVICE CORRECTION."
Customers were referred to BD's recall notice, which provides advice for users and instructions to mitigate hub breakage issues. Product return is not required.
If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Root Cause
Process change control
Distribution Pattern
US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.