DePuy Mitek LATARJET EXPERIENCE Top Hat Tap Product Code: 288203 Product Usage: The Latarjet Cortical Screw set is intended to treat recurrent shoulder instability by supporting the anteroinferior glenoid with a bony graft

FDA Recall Z-2618-2017 · Posted None

Recall StatusTerminated
Product CodeHWC
Date Initiated2017-05-12
Date Terminated2018-10-03
Product Quantity153 units

Reason for Recall

Combo Screw Driver (Product Code 288211) tip has the increased potential to break intra-operatively when being used at an angle off-axis to screw

Firm Action

DePuy Mitek issued recall letter dated May 12, 2017 advising of the problem and requesting return of the product. A Response Form is to be completed and returned. Questions: contact your DePuy Synthes Mitek Sales Consultant or Carolyn Somerville, DePuy Mitek, Inc. (via telephone: 508.828.3647 Fax: 508.828.3762 or via email: [email protected]).

Root Cause

Component design/selection

Distribution Pattern

US: AZ, CA, CO, GA, MA, MI, OH, WA Foreign: Austria, Australia, Belgium, France, Germany, Spain, Switzerland, Netherlands, Poland, UK

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.