Omega Medical Imaging - 2 over 2 Elevating Monitor Suspension pillar/column accessory Device accessory that elevates monitor for healthcare professionals.

FDA Recall Z-2621-2017 · Posted 2017-06-19

Recall StatusTerminated
Product CodeIWE
Date Initiated2017-03-23
Date Terminated2019-12-11
Product Quantity6 units

Reason for Recall

Elevating Monitor Suspension column dropping and will not rise back up.

Firm Action

The firm, OMEGA MEDICAL IMAGING, sent a "Field Safety Advisory Notice" dated 3/23/2017 to consignees. The notice described the product, problem and actions to be taken. The consignees were instructed to immediately examine the product and confirm correct product information; immediately discontinue use; move monitor suspension away from staff personnel and patients; and advising consignees to take photos of affected product and the P/N label. Photos of product labels are to be sent to [email protected] at Omega. Product is to be corrected. If there are any questions or concerns, please feel free to contact us at 407-323-9400 Ext 203 or email: omegamedicalimaging.com or [email protected].

Root Cause

Mixed-up of materials/components

Distribution Pattern

Worldwide Distribution -US Distribution to states of: IN, OH, PA, and MT; and country of: Pakistan.