Terminated

Interplant Radiation Treatment Planning System, Interplant Release 3.4.0 Used when planning and performing prostate brachytherapy implant procedure. It provides ultrasound probe alignment and radioactive seed implantation.

FDA Recall Z-2622-2011 · Posted 2011-06-23

Recall StatusTerminated
Product CodeMUJ
Date Initiated2008-12-01
Date Terminated2013-04-04
Product Quantity41

Reason for Recall

Interplant Software: If the trajectory needle with a non-zero retraction is adjusted in any of the planning views (Transverse, 3D) and the program is closed and reopened, the Needle Definitions page in the Dose Plan Summary (Plan View) displays the needle with a 0.00 retraction even when the needle is displayed correctly in the planning view. If the Eliminate Leading Spacers function is applied t

Firm Action

Computerized Medical Systems, Inc. sent a "CUSTOMER ADVISORY" letter dated December 1, 2008, to all affected customers. The letter identified the product, the problem, and the actions to be taken. Customers were instructed to discontinue use of the product. There is no workaround at this time. This issue was resolved in Interplant Release 3.4.1 software. All possibly affected users have been shipped the Interplant Release 3.4.1 software in May 2010. If you have any questions, please call 314-993-0003 or toll free 800-878-4267.

Root Cause

Software design

Distribution Pattern

Worldwide Distribution -- USA (nationwide) including states of: CA, KS, KY, MA, MD, ME, MO, NC, NE, NY, OH, PA, TX and VA and countries of: Australia, Brazil, Chile, China, Germany and Japan.