Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

Product Name: Salivary Testosterone ELISA Model/Catalog Number: SLV-3013 Product Description: An enzyme immunoassay for the quantitative in vitro diagnostic measurement of free active testosterone in saliva.

FDA Recall Z-2622-2024 · Posted 2024-08-15

Recall StatusOpen, Classified
Product CodeCDZ
Date Initiated2024-07-05
Product Quantity257 units

Reason for Recall

Invalid Controls. The Low Control is out of range; too high which causes invalid runs that would result in repeating the runs and delays for reporting results to patients.

Firm Action

On July 8, 2024 Urgent Field Safety Notification letters were sent to customers. Action to be taken by Distributors and End-users: Please check your inventory for the respective lots 66K014 and 66K123. and stop using these kits immediately and promptly quarantine them. If you faced this issue of failed controls, kindly report with daya, and request a replacement for the kits used that were affected. Also, all unused kits still in your stock of the affected lot(s) will also be replaced. Immediately reply to email: [email protected] to confirm this notice and your compliance. Please indicate in the email the quantity of kits used which failed and quantity of stock remaining. We sincerely apologize for the inconvenience this may have caused you and/or your customer(s).

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Nationwide distribution in the states of CA, IL, MN, PA, UT, VA.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →