BeneVision N1 Patient Monitor (N1). Part Numbers: 6660E-PA00017, 6660E-PA00014, 6660E-PA00012, 6660E-PA00011, 6660E-PA00018, 6660E-PA00015, 6660E-PA00020, Q6660E-PA00011, 6660E-PA00021, 6660E-PA00016, 6660E-PA00019, Q6660E-PA00018, 6660E-PA00010, Q6660E-PA00010, Q6660E-PA00021, Q6660E-PA00012, 6660E-PA00030, 6660E-PA00041, 6660E-PA00035, 6660E-PA00032, 6660E-PA00031. Software Versions: 1.04.00.01, 01.05.00.01, 01.06.00.01, and 01.06.01.01 The patient monitor is intended for monitoring, displaying, reviewing, storing, alarming and transferring of multiple physiological parameters.

FDA Recall Z-2622-2025 · Posted 2025-09-25

Recall StatusOpen, Classified
Product CodeMHX
Date Initiated2025-08-15
Product Quantity2,278 units

Reason for Recall

Potential for activation of an abnormal alarm pause.

Firm Action

Mindray contacted consignees via letter on about 08/15/2025. consignees were instructed that they may continue to use the BeneVision N1 patient monitor as a multi-parameter module with Mindray host monitors. Per the additional guidance, it is recommended to confirm that the remaining alarm time counter is actively displayed and follows the existing N1 Operator's Manual: Instructions for when the Alarm Pause is Activated. An image was also included in the notification letter showing the correctly displayed Alarm Pause notification and timer. Consignees were also requested to make all applicable clinical staff aware of the issue and to complete and return the Acknowledgement and Receipt form. Mindray's Care team will deploy a software update to correct the issue. Distributors were advised to contact their direct customers and notify them directly.

Root Cause

Software design

Distribution Pattern

Worldwide - Nationwide distribution and the country of Canada.