Palindrome Chronic Catheter Kit Symmetrical Tip14.5 Fr/Ch(4.8mm)x23cm

FDA Recall Z-2632-2023 · Posted 2023-09-25

Recalling FirmCovidien LP
Recall StatusOpen, Classified
Product CodeMSD
Date Initiated2023-08-22
Product Quantity3050 devices

Reason for Recall

Some catheter kits labeled as 23 cm implant length incorrectly included catheters of 28 cm implant length; the actual catheter implant length is identifiable based upon the correct labeling on the catheter body.

Firm Action

An URGENT: MEDICAL DEVICE RECALL notification letter was sent to customers on 8/22/23. Required Actions: 1. Immediately quarantine and discontinue use of all unused affected lots of the Palindrome" 14.5 Fr/Ch (4.8 mm) x 23 cm Chronic Catheter Kits. Please note: This recall does not include Palindrome Precision Catheter Kits. 2. Complete the online Customer Confirmation Form using the QR code located at the top of the first page of this letter or by using the link below. We request your response within 3 business days of receiving this notice. Replying promptly will confirm your receipt of this notification and prevent you from receiving additional notices. Customer Confirmation Link: https://www.novasyte.com/medtronic/fa1355 Note: Please complete the online Customer Confirmation Form even if you DO NOT have affected product. 3. If you have affected product, it must be returned for processing. Upon completion of the Customer Confirmation Form, Customer Service will contact you with an RGA number to return all unused affected Palindrome" 14.5 Fr/Ch (4.8 mm) x 23 cm Chronic Catheter Kits from your inventory to Medtronic. 4. If purchased from a distributor, contact your distributor directly to arrange for the return of the product back to your distributor. 5. This notice should be passed on to those who need to be aware within your organization or to any organization including, but not limited to, nephrologists, intensivists, implanting and managing physicians, renal nurses, critical care nurses, or other dialysis staff where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records. Medtronic has partnered with Novasyte, an IQVIA company, to assist in this recall. For assistance regarding online response processing or product return for this recall, please contact Novasyte using the information below: " Phone number: (855) 566-3445 " Email: [email protected] We regret any inconvenience this may cause. We a

Root Cause

Under Investigation by firm

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Brazil, China, Czech Republic, Finland, France, Germany, Italy, Netherlands, Norway, Poland, Portugal, Puerto Rico, Spain, Thailand, United Kingdom.