SUN DASH RADIUS 252 PN 8000809705. The device is intended to be used for the tanning of human skin.
FDA Recall Z-2635-2014 · Posted 2014-09-12
Advertisement
Recalling FirmJK Products & Services, Inc
Recall StatusTerminated
Product CodeLEJ
Date Initiated2014-07-25
Date Terminated2017-01-13
Product Quantity1948 units
Reason for Recall
The integrity of the wires inside main electrical chord become compromised from excessive flexing (during the door open/close process).
Firm Action
The owners of the affected device will be notified via certified letter. After the recalling firm confirms ownership, a technician will be dispatched to install a door clamp and door stop, and inspect the condition of the main electrical chord.
Root Cause
Device Design
Distribution Pattern
Nationwide Distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.