Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.

FDA Recall Z-2635-2023 · Posted 2023-09-25

Recall StatusOpen, Classified
Product CodeFRN
Date Initiated2022-04-22
Product Quantity1344 pumps

Reason for Recall

Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the Intermediate Pumping Chamber (IPC) of the administration set

Firm Action

Ivenix initiated contact on 4/22/22/ via phone communicating the contents of the customer letter which was then emailed to each customer. Letter states reason for recall, health risk and action to take: Please note, the specified start-up flow accuracy of up to +8% is maintained throughout the extended start-up time; and the steady-state flow accuracy of +l-SYo of the LVP is not affected with this condition. There is no corrective action you need to take at this time. We believe that this condition is low risk in that the affected flow rate range is limited, the start'up flow rate accuracy is unaffected and still within specifications, and the clinical practice of monitoring patients while administering high alert drugs mitigates the risk. Ivenix is working on a software update to correct this condition. It will be available in approximately 60 days. We will inform you when that update is ready to install. If you have additional questions, please contact your Ivenix representative or support below. Customer support - email Customer support - phone [email protected] I-978-775-8TAA

Root Cause

Device Design

Distribution Pattern

US Nationwide distribution in the states of NJ, WI.