Terminated

PTS Panels Test Strips, CHOL +HDL+GLU Panel for CardioChek PA brand analyzers, 15 tests 1 MEMo Chip, manufacturer Polymer Technology Systems, Inc., 7736 Zionsville Rd., Indianapolis, IN 46268.

FDA Recall Z-2641-2010 · Posted 2010-09-30

Recall StatusTerminated
Product CodeNBW
Date Initiated2010-07-06
Date Terminated2011-05-11
Product Quantity5805

Reason for Recall

The test strips appear to under-recover Cholesterol, HDL and glucose.

Firm Action

Distributors and direct customers were notified by letter on July 2, 2010. They were told to discontinue use and dispose of the affected lot of test strips. The firm also asked customers to complete an acknowledgement and confirmation document .

Root Cause

Other

Distribution Pattern

US accounts in: CA, MA, FL, NY, KY, VA, UT, MN, OH, and MD. Direct foreign accounts in: Belgium, England, Sweden, and Italy.