Masimo SPO2 Sensor, ear clip-adult (Merge Part Number: HW-HEMO-00047), Product Usage: Merge Hemodynamics monitors, measures, and records physiologic data from a human patient undergoing a cardiac catheterization procedure. The Monitoring System is for the monitoring of vital parameters including ECG, SpO2, invasive blood pressure, temperature, NIBP and C02, and for the evaluation of resting ECG, arrhythmias, ST-segments and cardiac output.

FDA Recall Z-2641-2017 · Posted None

Recall StatusTerminated
Product CodeDQA
Date Initiated2016-04-04
Date Terminated2019-10-15
Product Quantity2 sensors

Reason for Recall

Under certain conditions, the cables used with the Hemo system during a patient procedure could stop displaying the patient SpO2 causing a slight delay in patient care.

Firm Action

Merge sent an Urgent: Medical Device Recall letter dated March 28, 2016 via e-mail the week of 4/4/2016. If no email address was available, the letters were issued via certified mail. A second letter dated July 11, 2016 was issued to non-responders. The letter identified the affected product problem and actions to be taken.

Root Cause

Component change control

Distribution Pattern

Distribution was nationwide. There was also government and military distribution. There was no foreign distribution.